MLT-100V Non-destructive Micro-Leakage Tester

MLT-V100 Non-destructive Micro-Leakage Tester aka. Container Closure Integrity Tester (CCIT), employs non-destructive testing technology to meet the requirements of both vacuum decay and pressure decay testing methods. It is a highly versatile instrument, used for the professional testing of micro-leakage in a wide range of containers, including ampoules, vials, pre-filled syringes, bottles, non-porous pouches, blow fill seal strips medical devices.
Product Introduction

MLT-V100 Non-destructive Micro-Leakage Tester aka. Container Closure Integrity Tester (CCIT), employs non-destructive testing technology to meet the requirements of both vacuum decay and pressure decay testing methods. It is a highly versatile instrument, used for the professional testing of micro-leakage in a wide range of containers, including ampoules, vials, pre-filled syringes, bottles, non-porous pouches, blow fill seal strips medical devices. 


Product Features


● Built-in 10-inch touch-screen computer (External computer is optional)

● Single sample testing process can be completed within 15 seconds

● Micro-Leakage aperture (≥1μm) can be detected

● The test chamber and the main instrument are arranged separately, and the test chamber is suitable for more than 5 kinds of samples

● The test chamber is made of aluminum alloy or stainless steel, with pneumatic clamping system

● Built-in flow meter, easy to calibrate

● Three calibration methods for zero point, leakage aperture and flow

● Three judgment modes for test results: pressure decay, leakage aperture and leakage flow

● Leakage flow error ≤0.1ccm, Vacuum resolution ≤1pa/0.01mbar/0.0001psi, Multiple units are available kpa/mbar/pa/psi

● Suitable for various flexible and rigid samples including vials, infusion bags, packages for contact lenses, milk powder cans and electronic accessories, etc. with vacuum decay or pressure decay method


Test Standards


This instruments conforms to multiple international standards ASTM F2338: Standard Test Method for

Nondestructive Detection of Leaks in Packages by Vacuum Decay Method.


Technical Specification


ParameterMLT-V100MLT-V100 (T)MLT-V100 (P)
Test MethodVacuum DecayVacuum Decay/Pressure Decay
Measurement & ControlMicro-computer Control/Dual Sensors
Vacuum SystemVacuum Pump
Air SourceN/AExternal Air Source
UnitkPa/mbar/psi/Pa
Pressure Resolution1 Pa0.1 Pa
Test Range-99~0kPa-99~400kPa
Measuring Range≥0.01ccm (about1μm )
Test Accuracy±0.2μm (≥3μm )
Test Accuracy±0.3μm (<3μm )
Main Control ChipSTM Microcomputer Chip
Acquisition Frequency50~200Hz
Operation System7”LCD touch screen10.1”LCD touch screenComputer software
Data Storage600 GroupsUnlimited
Data CommunicationUSB/RS232/485
Body MaterialEngineering Plastics/Stainless Steel
Instrument Dimension320(L)×480(W)×320(H)mm340(L)×480(W)×320(H)mm340(L)×470(W)×320(H)mm
Instrument Weight20KG